Single-phase core transformers for medical rooms are manufactured according to DIN EN 61558-2-15 ( VDE 0570-2-15 ). They have separate windings with reinforced insulation.
They must not be rated smaller than 3.15 kVA and not larger than 8 kVA for IT systems.
The transformers are available in both vertical and horizontal versions.
Note: med. used transformers must not be protected against overload. They should only be protected against short circuits!
Characteristic properties of transformers for medical rooms: